High Flow Nasal Cannula: Be Careful Out There

As the saying goes the devil is in the details.  For some years now many centres worldwide have been publishing trials pertaining to high flow nasal cannulae (HFNC) particularly as a weaning strategy for extubation.  The appeal is no doubt partly in the simplicity of the system and the perception that it is less invasive than CPAP.  Add to this that many centres have found less nasal breakdown with the implementation of HFNC as standard care and you can see where the popularity for this device has come from.

This year a contact of mine Dominic Wilkinson@NeonatalEthics on twitter (if you don’t follow him I would advise having a look!) published the following cochrane review, High flow nasal cannula for respiratory support in preterm infants.  The review as with most cochrane systematic reviews is complete and comes to a variety of important conclusions based on 6 studies including 934 infants comparing use of HFNC to CPAP.

1.  No differences in the primary outcomes of death (typical RR 0.77, 95% CI 0.43 to 1.36; 5 studies, 896 infants) or CLD.

2.  After extubation to HFNC no difference in the rate of treatment failure (typical RR 1.21, 95% CI 0.95 to 1.55; 5 studies, 786 infants) or reintubation (typical RR 0.91, 95% CI 0.68 to 1.20; 6 studies, 934 infants).

3.  Infants randomised to HFNC had reduced nasal trauma (typical RR 0.64, 95% CI 0.51 to 0.79; typical risk difference (RD) -0.14, 95% CI -0.20 to -0.08; 4 studies, 645 infants).

4. Small reduction in the rate of pneumothorax (typical RR 0.35, 95% CI 0.11 to 1.06; typical RD -0.02, 95% CI -0.03 to -0.00; 5 studies 896 infants) in infants treated with HFNC but the RR crosses one so this may be a trend at best.

If one was to do a quick search for the evidence and found this review with these findings it would be very tempting to jump on the bandwagon.  Looking at the review a little closer though there is one line that I hope many do not miss and I was happy to see Dominic include it.

“Subgroup analysis found no difference in the rate of the primary outcomes between HFNC and CPAP in preterm infants in different gestational age subgroups, though there were only small numbers of extremely preterm and late preterm infants.”

In his conclusion he further states:

Further evidence is also required for evaluating the safety and efficacy of HFNC in extremely preterm and mildly preterm subgroups, and for comparing different HFNC devices.

With so few ELBW infants included and with these infants being at highest risk of mortality and BPD our centre has been reluctant to adopt this mode of respiratory support in the absence of solid evidence that it is equally effective to CPAP in these smallest infants.  A big thank you to our Respiratory Therapy Clinical Specialist for harping on this point over the years as the temptation to adopt has been strong as other centres turn to this strategy.

Might Not Be So Safe After All

Now do not take what I am about to say as a slight against my twitter friend.  The evidence to date points to exactly what he and his other coauthors concluded but with the release of an important paper in May by Taka DK et al, I believe caution is needed when it comes to our ELBW infants.

High Flow Nasal Cannula Use Is Associated with Increased Morbidity and Length of Hospitalization in Extremely Low Birth Weight Infants

This paper adds to the body of literature on the topic as it truly focuses on the outcome of infants < 1000g.  While this study is retrospective in nature it does cover a five year period and examines important outcomes of interest to this population.

The primary outcome in this case was death or BPD and whether HFNC was used alone or with CPAP, this was more frequent than when CPAP was used alone.  Other important findings were the need for multiple and longer courses of ventilation in those who received at least some HFNC.  In these times of overburdened health care systems with goals of improving patient flow, it is also worth noting that there was a significant prolongation of length of stay with use of HFNC or HFNC and CPAP.

One interesting observation was that the group that fared the worst across the board was the combination of CPAP and HFNC rather than HFNC alone.

CPAP (941) HFNC (333) HFNC +/- CPAP (1546)
CPAP d (median, IQR) 15(5-28) 7 (1-19)
HFNC d (median, IQR) 14(5-25) 13 (6-23)
HFNC +/- CPAP 15 (5-28) 14(5-25) 26 (14-39)
BPD or death % 50.40% 56.80% 61.50%
BPD % 42.20% 52.20% 59.00%
Multiple ventiation courses 51.10% 53.10% 64.70%
More than 3 vent courses 17.60% 21.00% 29.40%
Ventilator d (median, IQR) 18(5-42) 25 (6-52) 30 (10-58)

I believe the finding may be explained by the problem inherent with retrospective studies.  This is not a study in which patients were randomized to either CPAP, HFNC or CPAP w/HFNC.  If that were the case one would expect lung pathologies and severity of illness to even ou,t such that differences between groups might be explained by the difference in treatments.  In this study though we are looking though the rearview mirror so to speak.  How could we account for the combination being worse than the HFNC alone?  I suspect it relates to the severity of lung disease.  The babies who were placed on HFNC and did well on it might have had less severe chronic changes.  What might be said about those that had the combination?  Well, one could postulate that there might be some who were extubated to HFNC and collapsed needing escalation to CPAP and then failing that therapy were reintubated.  Another explanation could be those babies who were placed on CPAP after extubation and transitioned before their lungs were ready to HFNC may have failed and lost FRC thereby going back to CPAP and possibly intubation.  Exposure in either circumstance to HFNC would therefore put them at risk of further positive pressure ventilation and subsequent further lung injury.  The babies who could tolerate transition to HFNC without CPAP might be intermediary in their outcomes (as they were found to be) as they lost FRC but were able to tolerate it but consumed more calories leaving less for growth and repair of damaged tissue leading to prolonged need for support.

Either way, the use of HFNC was found to lead to worse outcomes and in the ELBW infant should be avoided as routine practice pending the results of a prospective RCT on the subject.

Is it a total ban though?

As with many treatments that one should not consider standard of care there may be some situations where there may be benefit.  The ELBW infant with nasal breakdown from CPAP that despite excellent nursing and RRT attention continues to demonstrate tissue damage is one patient that could be considered.  The cosmetic implications and potential for surgical correction at a later date would be one reason to consider a trial of HFNC but only in the patient that was close to being able to come off CPAP.  In the end I believe that if a ELBW infant needs non invasive pressure support then it should be with CPAP but as there saying goes there may be a right time and a place for even this modality.

 

 

11 thoughts on “High Flow Nasal Cannula: Be Careful Out There

  1. HFNC have crept into neonatal practice and totally replaced NCPAP in many neonatal units in the UK. There are obvious advantages to this therapy as alluded by you but which other therapy would you use in today’s age of EBM with hardly any RCTs in specific populations?

    I personally think that one of the big questions to answer in a large study is whether it increases duration of oxygen exposure and associated Co morbidities.

    Unfortunately some centres have moved to almost exclusive use of HFNC and have indirectly forced other units around them to not be able to use NCPAP.

    There may be no increased risk of mortality and morbidity but I don’t think we have the evidence to confidently say that NCPAP is part of the past and HFNC is the default mode for ELBW.

    Like

    1. What about bubble cpap with the ram?? I have many elbw children as young 25 weeks on the system. We have had very good luck with almost no tissue breakdown. I was not a fan at first due to leak and no Seal on the patient’s face. But it’s hard to argue with our outcomes

      Like

      1. The issue with the Ram in our experience is that the pressure being read is the same with the prongs on or off the face. Having said that we don’t use bubble cpap so perhaps the machine may play a role. Very comfortable for the patient though no question

        Like

      2. bubble with the RAM??? How’s the safety profile? What pressure is being delivered to your baby? How are you measuring it? cpap without a seal? Interesting…

        Like

  2. Would you change CPAP on fiO2 0,21 and PEEP 6 to HFNC with fiO2 0.25-30 in a 1000g 27w and actual coŕrected age of 31w, with compensated respieatory acidosis. For me it doesn,t make sense, but i would like to hear your opinion

    Like

    1. I don’t use HFNC at all so I can’t say what I would do from experience. Based on what I know and given our own units practice to allow feeding while on CPAP I would have to say I would go for the use of CPAP if I needed a reliable distending pressure

      Like

      1. Perhaps device and Flow Rates matter???? A Randomized Controlled Trial to Compare Heated Humidified High-Flow
        Nasal Cannulae with Nasal Continuous Positive Airway Pressure
        Postextubation in Premature Infants
        Clare L. Collins, MBChB, FRACP///// A total of 132 ventilated infants younger than 32 weeks’ gestation were randomized to receive either
        HHHFNC (n = 67) or NCPAP (n = 65). Extubation failure occurred in 15 (22%) of the HHHFNC group compared
        with 22 (34%) of the NCPAP group. There was no difference in the number of infants reintubated in the first
        week. Treatment with HHHFNC reduced the nasal trauma score 3.1 (SD 7.2) versus NCPAP 11.8 (SD 10.7),